User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Lab-Chron 1402:Timer DATE: 31-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 127 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 5335A STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P U F W 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following additional equipment is required for 1.004 DISP this procedure: 1.004 DISP Timer Calibration Fixture 1.005 DISP Connect the Stanford Research FS700 rear panel 10MHz 1.005 DISP OUTPUT to the TIME BASE input on the 5335A counter. 1.006 STD Stanford Research FS700 1.007 DISP Connect the UUT AC plug into the 1.007 DISP appropriate 115VAC receptacle. 1.008 OPBR Is the UUT counting? 1.009 JMPT 1.013 1.010 DISP Press the UUT start/stop switch. 1.011 OPBR Is the UUT counting? 1.012 JMPF 1.007 1.013 DISP Disconnect the UUT from the AC outlet. 1.014 DISP Connect the UUT AC plug to the Timer Calibration Fixture 1.014 DISP "TIMER" AC receptacle. 1.015 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 1.015 DISP position. 1.016 DISP Plug Timer Calibration Fixture AC plug into the 1.016 DISP appropriate 115VAC receptacle. 1.017 DISP Rotate the UUT thumbwheels in a downward direction 1.017 DISP for a reading of 00000. 1.018 DISP Connect the Timer Calibration Fixture as follows: 1.018 DISP TO COUNTER ........................ 5335A INPUT A 1.018 DISP VOLTAGE IN ........................ 5700A OUTPUT 1.019 HEAD {* Timer Calibration *} 1.020 JMP 2.005 1.021 EVAL 2.001 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 2.001 DISP position. 2.002 DISP Rotate the UUT thumbwheels in a downward direction 2.002 DISP for a reading of 00000. 2.003 IEEE [@5335]RE 2.004 JMP 2.009 2.005 IEEE [@5335][TERM LF] 2.006 IEEE [@5335]IN[D2000] 2.007 IEEE [@5335]FN2, AF1, CO1, TR1, AS1, AT1.25, BT4.75; 2.008 IEEE [@5335]WA1; 2.009 5700 5.00V S 2W 2.010 DISP Set the Fixture's OPEN/SHORT switch to the SHORT 2.010 DISP position. 2.011 OPBR Is the counter's GATE light on? 2.012 JMPF 2.001 2.013 HEAD TEST RUNNING 2.014 DISP When the UUT's display reads 10, set the Fixture's 2.014 DISP OPEN/SHORT switch to the OPEN position. 2.015 HEAD * Timer Calibration * 2.016 ACC T 0.001% 2.017 IEEE [@5335][I] 2.018 MEME 2.019 MEMI Use keyboard to enter actual UUT reading in Seconds: 2.020 MEME 2.021 MEMC T 1.000U #! Test Tol 1, Sys Tol 0.0001, TUR 10000.000 (>= 4.00). 3.001 DISP Rotate the UUT thumbwheel in a downward direction 3.001 DISP until the UUT reads all zeros. 3.002 JMP 4.003 3.003 EVAL 4.001 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 4.001 DISP position. 4.002 DISP Rotate the UUT thumbwheel in a downward direction 4.002 DISP until the UUT reads all zeros. 4.003 IEEE [@5335]RE;[D1000] 4.004 5700 5.00V S 2W 4.005 DISP Set the Fixture's OPEN/SHORT switch to the SHORT 4.005 DISP position. 4.006 OPBR Is the counter's GATE light on? 4.007 JMPF 4.001 4.008 HEAD [27][91]7m[27][91]5mTEST RUNNING 4.009 DISP When the UUT's display reads 100, set the Fixture's 4.009 DISP OPEN/SHORT switch to the OPEN position. 4.010 HEAD * Timer Calibration * 4.011 ACC T 0.001% 4.012 IEEE [@5335][I] 4.013 MEME 4.014 MEMI Use keyboard to enter actual UUT reading in Seconds: 4.015 MEME 4.016 MEMC T 1.000U #! Test Tol 1, Sys Tol 0.001, TUR 1000.000 (>= 4.00). 5.001 DISP Rotate the UUT thumbwheel in a downward direction 5.001 DISP until the UUT reads all zeros. 5.002 JMP 6.003 5.003 EVAL 6.001 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 6.001 DISP position. 6.002 DISP Rotate the UUT thumbwheel in a downward direction 6.002 DISP until the UUT reads all zeros. 6.003 IEEE [@5335]RE;[D1000] 6.004 5700 5.00V S 2W 6.005 DISP Set the Fixture's OPEN/SHORT switch to the SHORT 6.005 DISP position. 6.006 OPBR Is the counter's GATE light on? 6.007 JMPF 6.001 6.008 HEAD TEST RUNNING 6.009 DISP When the UUT's display reads 1000, set the Fixture's 6.009 DISP OPEN/SHORT switch to the OPEN position. 6.010 HEAD * Timer Calibration * 6.011 ACC 1000.00T 0.001% 6.012 IEEE [@5335][I] 6.013 MEME 6.014 MEMI Use keyboard to enter actual UUT reading in Seconds: 6.015 MEME 6.016 MEMC T 1.000U #! Test Tol 1, Sys Tol 0.01, TUR 100.000 (>= 4.00). 7.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.